9.4
Lid Open/Close Status Is Not Detected
LCD cover open/close indicator does not
function properly.
Open the case assembly (S2) and place it to the side.
Replace the limit switch (C6).
Replace the power board assembly (B4).
Replace the main board assembly (B2).
DBS-SH-e-1210
No
Is the right hinge (H13)
pressing down the limit
switch (C6) lever
properly?
Yes
Is the limit switch
(C6) normal?
No
Readjust the right hinge (H13) spacing.
Readjust the limit switch (C6) attachment
position.
Yes
Replace the limit switch (C6).
63