Service Menu - KERN DBS Service Manual

Electronic moisture analyzer
Table of Contents

Advertisement

1.2

Service Menu

The service mode configuration (menu configuration diagram) is shown below.
User Mode
Password Input
External Input
"@SVC=321"
"321"
Service Mode
END
Error Log Display (1)
Model Name Display
Factory Settings Display
Software Version Display
Span Temperature
Coefficient Display
(after aging)
Zero Drift Display
(after aging)
WAD Display
TAD Display
OAD Display
LAD Display
WG Display
DBS-SH-e-1210
Lights up during service mode
[↑ ]
r
E o Log D sp a
Erro
rr
play (3)
Error Log Display (2)
r Log
s
* [← ] and [→ ] keys can be used to view 5 errors.
[↓ ]
Only the most recent error is displayed.
* Displays items selected in factory settings.
Span Temperature
Coefficient Display
(before aging)
Zero Drift Display
(before aging)
* Display Example
rr r
E o Log D sp a
i l y (1)
rr i r l y
4) i
y (5
E o Log D sp a
)
l
(
[Normal Range (guideline)]
Span after aging
Span before aging 0 to ±10.000 ppm/°C
Zero after aging
: [← ][→ ]
Zero before aging
: [↑ ][↓ ]
: Next [ENTER]
WAD
Back [← ]
TAD (10 to 30 °C)
OAD
: Next [ENTER]
LAD
WG
THAD
TT
ST
*1) Loading pan(S14)( t0.3 Aluminum)
0 to ±3.000 ppm/°C
0 to ±0.002 g/°C
0 to ±0.015 g/°C
Zero(*1)
Span (60 g)
300,000 to
Approx.
800,000
1,550,000
26,000 to 40,000
±10.000
Weighing
capacity (g)
(g)
At TT = 25 °C:
approx. 320 to 480
At TT = 100 °C:
approx. 150 to 220
Room temperature ±10 °C
Room temperature ±10 °C
6