9.7

ERR.124 Occurred

ERR.124 occurred.
Is warm-up time
Yes
direct sunlight?
No
Exposed to air
No
Does not reach
temperature
specified when
Yes
See 9.2 Reproducibility of Weight
Measurements is Poor.
66
No
adequate?
Exposed to
Yes
Yes
flow from a
heater or
cooler?
measuring
moisture?
Warm up.
Change the installation location of the
moisture analyzer.
Change the installation location of the
moisture analyzer.
No
See 9.6 Does Not Reach
Temperature Setting.