40
information@ITM.com
G.2.3 Draught
The
function is only available when a fl ue gas probe is connected.
Draught
Do not measure for longer than 5 min, as the drift of the pressure sensor means that
the readings could be outside the tolerance limits.
Calling up the function:
OK
Measurements
Measuring:
1 Start measuring:
Start
.
- Draught zeroing (5 s).
2 Position the fl ue gas probe in the centre of the fl ow (area of the highest fl ue gas
temperature). The display showing the maximum measured fl ue gas temperature (FT)
helps when positioning the probe.
- The reading is displayed.
3 Stop measuring
Stop
.
- The reading is recorded.
Option:
To print the reading:
Print
4 To copy the reading to the
- The
Measurements
menu is opened.
G.2.4 Smoke# /HCT
Calling up the function:
OK
Measurements
Recording smoke tester no. / smoke numbers / oil derivative with the smoke pump
and manual input:
The function is only available if the chosen fuel is an oil.
1
Sm. tester no.
Change
Enter the tester number
2
Change
Enter the value
Smoke # 1
3 Repeat step 2 for the other smoke # and the oil derivative accordingly.
www.
OK
.
Draught
.
menu:
OK
.
Flue gas
OK
.
Smoke#/HCT
OK
OK
.
Testo-Direct
.