Set Current Mode

Define either a 0-to-20 or 4-to-20 mA analog output range for the selected
analog output port. This output is dependent on both the range of read-
ings you define and on the function you track on the analog output.
Setting the current mode
1.
Select an analog output port.
2.
From the Analog Range key, select Set Current Mode.
3.
Select 4-to-20 mA Mode or 0-to-20 mA Mode. Default selection is 4-to-
20 mA.
2.
Select 4 to 20 mA Mode or 0 to 20 mA Mode.
Out # 1 will TRACK
during Calibration
or
Out # 1 will HOLD
during Calibration
Hold during Verify
or
Track during Verify
Out # 1 will TRACK
during Verify
or
Out # 1 will HOLD
during Verify