Dual Link Format
Verification Test
WVR5200 Waveform Rasterizer Specifications and Performance Verification
number in each box as required. Accept the changes by moving the
highlight to the Accept box and press SEL.
4. Enable the instrument Web access from the Configuration menu by selecting
Network Settings > Web Enable > On.
5. Open a Web browser on the computer.
6. From the computer, enter the IP address of the instrument into the Web
browser address line (for example, http://192.168.1.1). This is the address that
appears in the field CONFIG > Network Settings > IP Address. Note that
leading zeros are deleted in the address line.
7. You should see a Web page titled "Tektronix WFM5200 Interface" or
something similar. This means the Ethernet function is working.
8. Record Pass or Fail for Ethernet Functionality in the test record.
1. Set up a dual link signal generator (TG700 with HDLG7 module). Connect
the Link A output on the module to the SDI 1A input of the instrument, and
the Link B output to the SDI 1B input of the instrument.
2. Perform the following steps to set the HDLG7 module format to 1080 59.94i:
a. Press the MODULE button until HDLG7 appears.
b. On the signal generator, press the FORMAT button until 1080 59.94i
appears and then press the ENTER button.
c. Press the CANCEL button.
d. Press the up/down arrows until Sample Structure/Depth appears.
e. Press the left/right arrows until "YCbCr+A 4:4:4 10 bits" appears.
f.
Press the ENTER button.
g. Press the COLOR BAR button until 100% color bars appears.
3. Restore the Factory Preset. (See page 32, Restore the Factory Presets.)
4. Press the Display Select (Thumbnail) button and select tile 4.
5. Press and hold the STATUS button.
6. Select Display Type > Video Session.
7. Press the Display Select (Thumbnail) button and select tile 2.
8. Press and hold the VECTOR button to display the Vector menu.
9. Select Display Type > Lightning.
10. Press the VECTOR button to close the menu.
Incoming Inspection
45
Need help?
Do you have a question about the WVR5200 and is the answer not in the manual?
Questions and answers