4.2.11 Adapter Function Measurement

4.2.11.1 In DC mode
Measurement Range (1)
Specified Measurement Range (2)
Accuracy
Resolution
Input Impedance
4.2.11.2 In AC mode
Measurement Range (1)
Specified Measurement
Range (2)
Accuracy
Resolution
Input Impedance
Note (1) The basic display is 10,000 counts. The position of the decimal count
and the display of multiples (m and k) depend on the programming of
the scale factor.
Note (2) - The max. bandwidth is 1kHz.
Specific specifications in MAX/MIN mode (from 10Hz to 1kHz):
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.
4.3

ENVIRONMENTAL CONDITIONS

Conditions
Temperature
Relative Humidity (RH)
measurement range
100.0 to 999.9mV
In DC, the display indicates "+OL" above +9999 counts and "-
OL" below -9999 counts. The "-" and "+" signs are managed
(polarity).
In AC, the display indicates "OL" above 9999 counts.
Operating
-4° to +131°F
(-20° to +55°C)
≤90% up to 131°F (55°C)
0.0 to 999.9mV
0 to 100% of the measurement range
0.1mV
5.0 to 999.9mV
1 to 100% of the
5.0 to 99.9mV
± (1% R + 10cts)
± (1% R +3cts)
0.1mV
36
1.00 to 9.99V
1% R +3cts
10mV
10MΩ
1.00 to 9.99V
0 to 100% of the
measurement range
1% R +3cts
10mV
10MΩ
Storage
-40° to +158°F
(-40° to +70°C)
≤90% up to 158°F (70°C)