3.13 ADAPTER FUNCTION MEASUREMENT

This function makes it possible to connect any adapter or sensor that converts a
measured electrical or physical quantity into a voltage (either DC or AC). A direct,
immediate reading is obtained without applying a conversion factor.
The measurement mode, AC or DC (the default), must be chosen manually using
the yellow button.
The scale factor for the adapter or sensor must be chosen in advance in the
instrument's
set-up
adapter/sensor sensitivities that allow a direct reading once the scale factor has
been chosen.
NOTE: The examples given in Amps are valid for other quantities:
humidity (%RH), illumination (lux), speed (m/s), flow, (GPM) etc.
1.
Connect the black lead to the COM terminal and the red lead to the "+"
terminal.
2.
Set the switch to
3.
Select the AC or DC mode.
4.
Connect the adapter according to its directions for use.
(§3.4.5).
Adapter/Sensor
Sensitivity
(examples in Amps)
10 mV/kA (0.01 mV/A)
100 mV/kA (0.1 mV/A)
1 mV/A
10 mV/A
100 mV/A
1000 mV/A (1 mV/mA)
10 mV/mA
100 mV/mA
.
The
table
below
Scale Factor to be
Programmed
30
indicates
the
10 k
100 k
1
10
100
1 m
10 m
100 m
various