Process Error Messages; Profibus Dp/Pa; List Of Process Error Messages - Endress+Hauser Proline Promag 50 Operating Instructions Manual

Electromagnetic flow measuring system profibus dp/pa
Hide thumbs Also See for Proline Promag 50:
Table of Contents

Advertisement

Troubleshooting
!
No.
Device status message
(local display)
P = Process error
"Fault message" error message type:
• If this message occurs, operation is immediately interrupted or stopped!
• Local display → A lightning symbol ($) flashes on the display
"Notice message" error message type:
• Normal operation continues despite this message!
• Local display → An exclamation mark (!) flashes on the display.
Error messages on the local display → See Table
401
P EMPTY PIPE
! # 401
461
P EPD ADJ. N. OK
! # 461
463
P FULL = EMPTY
$ # 463
98
9.3

Process error messages

Note!
Also observe the information on Page 63 and Page 99.
9.3.1
Displaying the device status on PROFIBUS DP/PA
Further information see Page 94 ff.
9.3.2

List of process error messages

PROFIBUS measured value status
0x50
UNCERT
sensor
AIN
convention
not accurate
(measured
value from
sensor
inaccurate)
0x40
UNCERT
non-specific
AIN
(uncertain
status)
0x40
UNCERT
non-specific
AIN
(uncertain
status)
.
Adv. diagnostics
message in
PROFIBUS
master
no limits
Empty pipe
detected
no limits
EPD adj. not
possible
no limits
EPD adj. wrong
Proline Promag 50 PROFIBUS DP/PA
Cause/remedy
Cause:
Alarm from empty pipe detection (EPD).
The measuring tube is only partially filled or empty.
Remedy:
1. Check the process conditions of the plant.
2. Fill the measuring tube.
Cause:
EPD adjustment not possible because the fluid's
conductivity is either too low or too high.
Remedy:
The EPD function cannot be used with fluids of this
nature.
Cause:
The EPD adjustment values for an empty pipe and full
pipe are identical, therefore incorrect.
Remedy:
Repeat EPD adjustment, observing procedure closely
→ Page 101 ff.