Nominal Diameter And Flow Rate - Endress+Hauser Proline Promag 50 Operating Instructions Manual

Electromagnetic flow measuring system profibus dp/pa
Hide thumbs Also See for Proline Promag 50:
Table of Contents

Advertisement

Installation
!
20
3.2.7

Nominal diameter and flow rate

The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum velocity of flow is 2...3 m/s. The velocity of flow (v), moreover, has to be matched to the
physical properties of the fluid:
• v < 2 m/s: for abrasive fluids such as potter's clay, lime milk, ore slurry, etc.
• v > 2 m/s: for fluids producing buildup such as wastewater sludge, etc.
Note!
Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor
→ Page 19.
Flow characteristics
Measuring device
min./max. full scale value (v ~ 0.3 or 10 m/s)
nominal diameter
[mm] [inch] Promag
1
2
/
"
H
12
5
4
/
"
H
32
5
8
/
"
H
16
15
½"
P, H
25
1"
W, P, H
32
1¼"
W, P
40
1½"
W, P, H
50
2"
W, P, H
65
2½"
W, P, H
80
3"
W, P, H
100
4"
W, P, H
145...4700 dm
125
5"
W, P
220...7500 dm
150
6"
W, P
200
8"
W, P
250
10"
W, P
300
12"
W, P
350
14"
W, P
110...3300 m
400
16"
W, P
140...4200 m
450
18"
W, P
180...5400 m
500
20"
W, P
220...6600 m
600
24"
W, P
310...9600 m
700
28"
W
420...13500 m
30"
W
480...15000 m
800
32"
W
550...18000 m
900
36"
W
690...22500 m
1000
40"
W
850...28000 m
-
42"
W
950...30000 m
1200
48"
W
1250...40000 m
54"
W
1550...50000 m
1400
W
1700...55000 m
-
60"
W
1950...60000 m
1600
W
2200...70000 m
-
66"
W
2500...80000 m
1800
72"
W
2800...90000 m
-
78"
W
3300...100000 m
2000
W
3400...110000 m
Recommended flowrate
SI unit
US unit
3
0.06...1.8 dm
/min
0.015...0.5 gal/min
3
0.25...7 dm
/min
0.07...2 gal/min
3
1...30 dm
/min
0.25...8 gal/min
3
4...100 dm
/min
1.0...27 gal/min
3
9...300 dm
/min
2.5...80 gal/min
3
15...500 dm
/min
4...130 gal/min
3
25...700 dm
/min
7...190 gal/min
3
35...1100 dm
/min
10...300 gal/min
3
60...2000 dm
/min
16...500 gal/min
3
90...3000 dm
/min
24...800 gal/min
3
/min
40...1250 gal/min
3
/min
60...1950 gal/min
3
20...600 m
/h
90...2650 gal/min
3
35...1100 m
/h
155...4850 gal/min
3
55...1700 m
/h
250...7500 gal/min
3
80...2400 m
/h
350...10600 gal/min
3
/h
500...15000 gal/min
3
/h
600...19000 gal/min
3
/h
800...24000 gal/min
3
/h
1000...30000 gal/min
3
/h
1400...44000 gal/min
3
/h
1900...60000 gal/min
3
/h
2150...67000 gal/min
3
/h
2450...80000 gal/min
3
/h
3100...100000 gal/min
3
/h
3800...125000 gal/min
3
/h
4200...135000 gal/min
3
/h
5500...175000 gal/min
3
/h
9...300 Mgal/d
3
/h
10...340 Mgal/d
3
/h
12...380 Mgal/d
3
/h
13...450 Mgal/d
3
/h
14...500 Mgal/d
3
/h
16...570 Mgal/d
3
/h
18...650 Mgal/d
3
/h
20...700 Mgal/d
Proline Promag 50 PROFIBUS DP/PA
Factory settings
Low flow cutoff (v ~ 0.04 m/s)
SI unit
3
0.01 dm
/min 0.002 gal/min
3
0.05 dm
/min 0.008 gal/min
3
0.1 dm
/min 0.025 gal/min
3
0.5 dm
/min
0.10 gal/min
3
1.0 dm
/min
0.25 gal/min
3
2.0 dm
/min
0.50 gal/min
3
3.0 dm
/min
0.75 gal/min
3
5.0 dm
/min
1.25 gal/min
3
8.0 dm
/min
2.0 gal/min
3
12.0 dm
/min
2.5 gal/min
3
20.0 dm
/min
4.0 gal/min
3
30.0 dm
/min
7.0 gal/min
3
2.5 m
/h
3
5.0 m
/h
3
7.5 m
/h
3
10 m
/h
3
15 m
/h
3
20 m
/h
3
25 m
/h
3
30 m
/h
120 gal/min
3
40 m
/h
180 gal/min
3
50 m
/h
210 gal/min
3
60 m
/h
270 gal/min
3
75 m
/h
300 gal/min
3
100 m
/h
360 gal/min
3
125 m
/h
480 gal/min
3
125 m
/h
600 gal/min
3
150 m
/h
600 gal/min
3
200 m
/h
1.3 Mgal/d
3
225 m
/h
1.3 Mgal/d
3
250 m
/h
1.3 Mgal/d
3
300 m
/h
1.7 Mgal/d
3
325 m
/h
2.2 Mgal/d
3
350 m
/h
2.6 Mgal/d
3
450 m
/h
3.0 Mgal/d
3
450 m
/h
3.0 Mgal/d
Endress+Hauser
US unit
12 gal/min
15 gal/min
30 gal/min
45 gal/min
60 gal/min
60 gal/min
90 gal/min

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents