EMI Receiver Instructions for Use
Add extreme frequency points: Click [Add]/[Insert], the limit line frequency point list will automatically add a line, pulse
The knob moves to the frequency or limit table that needs to be modified, input and modify through the number keys on the panel.
After editing, click [Save], and the system will automatically return to the limit line list interface.
5.1.3.2 Enabling limit lines
[Meas Setup]Enter the limit line list home page
Use the pulse knob to switch between enable and trace form, enable or cancel by [Enter] key, trace selection
The selection range is from 1 to 3, and the corresponding trace is selected according to the actual set parameters. After reset, the system automatically cancels all
There are limit lines.
5.2 Instrument Compensation Factor Editing
Compensation factors can be edited for testing accessories such as LISN, probe, antenna, limiter, attenuator, amplifier, etc.
Corresponding compensation is made for the insertion loss and frequency response to ensure accurate and reliable test results.
[Meas]→[Compensation],
Operation
steps:
enter
the
compensation
factor interface
44