EMI Receiver Instructions for Use
step
Resolution bandwidth 120kHz
Measurement time 100us
Attenuation
Preamp
Description of parameters in the above figure: start frequency 1kHz, stop frequency 300MHz, minimum level 0dBuV, level
Display range 100dB, continuous scanning. Specific parameter input is mainly by means of up and down direction keys and pulse rotation
button to switch and input. in,
Scan start: set the scan frequency at the start of the interval, and modify it by inputting the number keys
Scanning termination: set interval termination scanning frequency, number key input modification
Stepping: currently in automatic mode, without modification, it will be automatically assigned by the system
Resolution bandwidth: 6dB resolution bandwidth selection, switch by [Enter]
Measuring time: scanning time of stepping frequency, number key input modification
Attenuation: 0~39dB input range, 3dB step, can be modified by numeric key input, or by [Enter]
to switch
Pre-amp: Built-in 20dB amplifier switch, open when measuring small signals
2.1.3.4 Standard measurements
(1) EMI receiver system entry steps
Take the GJB151B CE102 test as an example:
[Mode]→[EMI]→[Meas]→[Standard]→[CE102]→[220V]
NOTE: Follow these steps for other standard inputs.
40kHz
9dB
off
21