EMI Receiver Instructions for Use
Open the network measurement function, press the arrow keys【⬆】【⬇】or number keys to change the output power, reference level,
Scan time and scan points.
—Press the [Normalization] key, and the display will display a straight line in the center of the screen, as shown in Figure 3,28.
The normalized line is at the center of the screen
If you want to expand the vertical display range, you can adjust the reference level soft key to adjust the normalized line to the top of the display
part, as shown in Figure 3,29.
The normalized line is at the top of the screen
2. Remove the adapter and replace the adapter with a filter to display the filter characteristics, as shown in Figure 3.30
shown.
Figure 3.28 Normalized measurement
adapter
Figure 3.29 Normalized measurement
29
adapter