Emi receiver
Table of Contents

Advertisement

EMI Receiver Instructions for Use
2.1.3 Common EMI Receiver Operations
2.2.3.1 Common parameter settings
Common parameters of the receiver include: start frequency, stop frequency, attenuation, preamplification, dynamic range, bandwidth,
Measurement time, scan mode, etc.
【Start Frequency】
Operation steps: [FREQ]→[Start Frequency], the default is 30MHz, after adjustment, the frequency spectrum displays the area
frequency
rate range auto-adjust
【Stop frequency】
Operation steps: [FREQ]→[Stop Frequency], the default is 1GHz, after adjustment, the spectrum displays the regional frequency
auto range adjustment
【attenuation】
Operation steps: [AMPT] → [Attenuation], the default is 9dB, the range of change is 0~39dB, 3dB step
【Pre-amplification】
Operation steps: [AMPT] → [Pre-amplification], the default is off, and the gain can be increased by 20dB after opening
【Dynamic Range】
The spectrum display area displays the range, which can display 100dB dynamic range by default, and the minimum level is 0dBuV.
Operation steps: [AMPT] → [level range], the default is 100dB, and the adjustable range is 10dB~200dB
[AMPT] → [Minimum level], default 0dBuV
-93dBuv~297dBuV
After the setting is completed, the dynamic range of the spectrum display area is automatically updated.
【bandwidth】