EMI Receiver Instructions for Use
5. After each measurement, first unplug the connecting cable between the monitoring terminal of LISN and the spectrum analyzer,
Then turn off the power of the spectrum analyzer, and finally turn off the power of the device under test.
6. Because there is a relatively large filter capacitor in the power supply impedance stabilization network, it cannot be used in leaky
on the electrical protection grid. If it is used in a general laboratory, it must be transformed by power isolation
The transformer feeds the mains impedance stabilization network.
After setting up the experimental test platform, follow the steps below to test:
1: After booting and warming up for 25 minutes, the system will automatically enter the EMI receiver system interface;
2: Test standard selection. PressÿMeasÿto select the EMC test standard, and use the softkeys on the interface to directly select
Select [Standard], select [GJB151A] in the standard list, select [CE102] in the national military standard list, according to
The DUT is required to select the [220V] reference line, and the test interface under this standard is as follows:
Figure 3.13 CE102 interface
25