ER series EMI receivers have network measurement function and are mainly used for amplitude-frequency characteristic measurement. can be measured
Insertion loss, isolation, and amplifier gain for passive networks such as filters, attenuators, and cabling
wait. For example, to measure the performance of the filter: connect the output of the signal tracking source to the input of the filter, and then connect the filter
Connect the output terminal of the oscilloscope to the input terminal of the spectrum analyzer to measure the filtering characteristics of the filter.
Measurement methods:
Setting parameters
—Press [Mode]key to select [Network Measurement]
—Press[SPAN]key to adjust the span to the required width (eg. 200MHz)
1. Use an adapter (short connector) to connect the tracking generator output cable to the spectrum input cable,