Line Impedance Stabilization Network (LISN) Instructions for Use WARNING
1. When using the power supply impedance to stabilize the network, the ground terminal on the rear panel of the instrument must be connected to the reference ground.
Reliable connection.
2. Only when the equipment under test (EUT) works normally, the monitoring of the power supply impedance stabilization network
Connect the terminal to the input terminal of the spectrum analyzer. In order to protect the safety of the test equipment, avoid the signal too
If it is too strong and burns out the receiver, a 20dB attenuator should be added to the receiver input.
3. When the monitoring terminal of the source impedance stabilization network (LISN) is connected to the receiver, do not
Power
off the
may cause overvoltage spikes and damage the receiver.
4. When the monitoring end of the power supply impedance stabilization network is not connected to the receiver, it should be connected to a monitoring end
50Ω
load.
Figure 3.20 CE102 test block diagram and specifications
Figure 3.11 Build CE102 test platform
Figure 3.12 Specific Field Test Connection Diagram