EMI Receiver Instructions for Use
Operation steps: [BW] → [Resolution Bandwidth] Receiver mode provides
10Hz/100Hz/200Hz/1kHz/9kHz/10kHz/100kHz/120kHz/1MHz total 9 file resolution
Bandwidth, you can cycle through the up and down arrow keys, and the menu area displays commonly used
200Hz/9kHz/120kHz/1MHz four-speed resolution bandwidth
【measure time】
Receiver measurement time range: 100us~100s
Operation steps: 【Meas】→【Measuring time】, the default is 10ms
【Scan mode】
Operation steps: [Sweep] → [List Sweep] / [Current Settings]
[List Scan] mode provides 10 independent measurement intervals, and each interval can be set for different measurements
Parameters, executed in interval order during measurement
【Current setting】Quick measurement, equivalent to an interval measurement
2.2.3.2 Fixed frequency measurement
Fixed frequency measurements are automatically activated when not scanning, and automatically hide after starting a scan or measurement
The fixed frequency is determined by the receiving frequency, and the maximum value (MAX), quasi-peak value (QPK),
Average value (AVG), effective value (RMS) four kinds of detection results display, each detection value has a maximum hold
Hold function in dBuV. As shown in the figure above, the maximum value (MAX) detection, the current real-time value is 47.56dBuV,
The historical maximum is 48.46dBuV. After the receiving frequency is modified, the maximum hold value is automatically reset.
19