EMI Receiver Instructions for Use
21.
2) Parameter setting
—PressÿModeÿkey
—Press [EMI Receiver]
— by [Standard]
- Press [Conduct]
—Press [CE102-1]
3) Total interference measurement
After the parameters are set, connect the detection terminal "L Output/N Output" of the front panel LISN to the frequency spectrum
After the scan is stable, observe whether the curve exceeds the standard limit, and adjust the frequency to exceed the standard limit.
On the target spectral line, at this time, the total amount of interference exceeds the standard by 22dB at 131kHz. As shown in Figure 4.8.
Figure 4.7 Connection Diagram of CE102 Conduction Test Example