Emerson Rosemount 702 Quick Installation Manual page 11

Wireless discrete transmitter
Hide thumbs Also See for Rosemount 702:
Table of Contents

Advertisement

Quick Installation Guide
00825-0100-4702, Rev BA
August 2009
Hazardous Locations Certificates
North American Certifications
Factory Mutual (FM) Approvals
I5 FM Intrinsically Safe, Non-Incendive and Dust Ignition-Proof
Intrinsically Safe for Class I/II/III, Division 1, Groups A, B, C, D, E, F, and G.
Zone Marking: Class I, Zone 0, AEx ia llC
Temperature Codes T4 (-50 °C <= T
Non-incendive for Class I, Division 2, Groups A, B, C, and D.
Dust Ignition-proof for Class II/III, Division 1, Groups E, F, and G.
Intrinsically Safe and non-incendive when installed in accordance with Rosemount
drawing 00702-1000.
For use with Rosemount SmartPower
Enclosure Type 4X / IP66 / IP67
Canadian Standards Association (CSA)
I6 CSA Intrinsically Safe
Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D.
Temp Code T3C
Enclosure Type 4X / IP66 / IP67
For use with Rosemount SmartPower Options P/N 753-9220-0001 only
Intrinsically Safe when installed per Rosemount drawing 00702-1020
European Certifications
I1 ATEX Intrinsic Safety
Certificate No.: BASEEFA07ATEX0239X
Ex ia IIC T4 (-60 °C <= T
IP66 / IP67
For use with Rosemount SmartPower ™ options P/N 753-9220-XXXX only
Special conditions for safe use (X)
The surface resistivity of the antenna is greater than 1 gigaohm. To avoid electrostatic
charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
1180
TABLE 2. Sensor Parameters
Sensor
U
= 6.6 V
o
I
= 26 mA
o
P
= 42.6 mW
o
C
= 10.9 uH
o
L
= 500 mH
o
<= 70 °C), T5 (-50 °C <= T
amb
®
Options P/N 753-9220-0001 only.
<= 70 °C), Ex ia IIC T5 (-60 °C <= T
amb
II 1G
Rosemount 702
<= 40 °C)
amb
<= 40 °C)
amb
11