Chapter 12: Error Messages
Graphical/Softkey Interface Messages - Unnumbered
Performance tool must be initialized
Cause: You tried to make a performance measurement when the Software
Performance Measurement Tool (SPMT) was not initialized.
Action: The Software Performance Measurement Tool (SPMT) must be initialized
before making performance measurements on your software. Use the
performance_measurement_initialize command to initialize the SPMT.
Performance tool not initialized
Cause: The Software Performance Measurement Tool (SPMT) has not been
initialized.
Action: To make accurate activity or duration measurements on current data, use
the performance_measurement_initialize command to initialize the SPMT before
running a performance measurement.
Question file missing or invalid
Cause: Some of the Softkey User Interface files are missing or are corrupted.
Action: Reinstall the host software and try starting the emulation session again.
Range crosses segment boundary
Cause: On a segment offset processor, an address range is specified that would
cross different segments.
Action: Break the memory command into multiple commands so that the address
ranges start and end in the same segment.
Read memory failed at <PHYSICAL ADDRESS> - store aborted
Cause: While storing memory from the emulator to a file, a read memory error
occurred.
Action: Use the display error_log command to view any errors. You may need to
modify the configuration and map memory before storing the file again.
Session aborted
Cause: This will only happen when running multiple emulation windows and a
fatal system error occurs.