Remove the filter case by using maintenance handle which is supplied with product.
Please note that circulating fluid may flow out from the case when removing filter case.
Discharge circulating fluid inside of filter case.Do not drop out filter element from inside of filter
case during this operation.
Filter case
After discharging circulating fluid, insert the element into the resin filter case. Mount the resin
filter case by hand tighten.
Mounting of the case should be performed by hand-tighten. When
the case is tightened excessively by tools or handle, the case might
crack or be damaged.
Make sure that the case is held manually by hand when removing
and mounting the case. If the case is dropped, it may result in
cracking or breakage.
6.
Use a cross slot screwdriver to install the filter panel.
(Recommended tightening torque:1.5N・m)
HRR Series
Chapter 8 Control, Inspection, Exchange and Cleaning
Maintenance handle
(Accessory )
Caution
HRX-OM-X071
8.3 Stop for a Long Time