Environmental Requirements - Keysight Technologies N5251A Series Manual

2-port and microwave network analyzer system
Table of Contents

Advertisement

System Installation
Site Preparation

Environmental Requirements

The environmental requirements shown below are characteristic for the system and are
based on the limitations of the N5227A network analyzer used.
Table 3-4 Environmental Requirements
Temperature
Operation
Storage
Measurement Calibration
Performance Verification
Relative Humidity
Pressure Altitude
System Heating and Cooling
Install air conditioning and heating, if necessary, to maintain the ambient temperature within the
appropriate range (as given in the table above). Air conditioning capacity must be consistent
with the BTU ratings given in
Required Conditions for Accuracy Enhanced Measurement
Accuracy-enhanced (error-corrected) measurements require the ambient temperature of the
N5251A to be maintained within ± 1
3-6
5 C to 40 C (41 F to 104 F)
-40 C to +65 C (-40 F to +158 F)
20 C to 26 C (68 F to 79 F)
Temperature must be within 1 C (1.8 F)
of the temperature at which the
measurement calibration was performed.
Type tested at 95%, +40 C (non-condensing)
Type tested 0 to 4600 meters (~15,000 feet)
Table 3-3 on page
3-5.
C of the ambient temperature at calibration.
PNA Series Microwave Network Analyzer System
Installation and Service Guide N5251-90001
N5251A

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5227aN5261aN5262a

Table of Contents