Example before Auto False Echo Suppression
①
②
③
④
Example after Auto False Echo Suppression
①
②
③
SITRANS LR250 with mA/HART
Operating Instructions, 10/2019, A5E32220602-AG
default TVT
false echo
material echo
sensor reference point
learned TVT
false echo
Auto False Echo Suppression Range
Appendix A: Technical reference
⑤
high calibration point = 0
⑥
obstruction at 1.3. m
⑦
material level at 3.2 m
④
default TVT
⑤
material echo
⑥
echo marker
A.2 Echo Processing
Auto False Echo
Suppression
Range set to 2 m
239