Parameter reference
After Auto False Echo Suppression
①
②
③
To set Auto False Echo Suppression via SIMATIC PDM:
Open the menu Device – Echo Profile Utilities and click on the tab Auto False Echo
Suppression. For more detailed instructions see Auto False Echo Suppression via SIMATIC
PDM (Page 80).
To set Auto False Echo Suppression via the handheld programmer:
Options
The learned TVT takes effect only at the next measurement.
a)
134
false echo
learned TVT
Auto False Echo Suppression
Range
OFF
*
ON
LEARN
④
material level
⑤
default TVT
⑥
echo marker
Default TVT will be used.
'Learned' TVT will be used.
'Learn' the TVT
.
a)
Operating Instructions, 10/2019, A5E32220602-AG
SITRANS LR250 with mA/HART