Symptom
Reading response
slow
Reads correctly but
occasionally reads
high when vessel is
not full
Level reading lower
than actual material
level
SITRANS LR250 with mA/HART
Operating Instructions, 10/2019, A5E32220602-AG
Cause
liquid surface vortexed
material filling
Fill Rate per Minute (2.4.2.) set-
ting incorrect
detecting close range echo
•
build up near top of vessel or
•
nozzle
nozzle problem
•
material is within Near Range
•
zone
multiple echoes processed
•
vessel near empty and low
•
dK material
Action
decrease Fill Rate per Minute (2.4.2.)
•
relocate device to side pipe
•
increase confidence threshold in Echo Threshold (2.8.4.3.)
•
Re-locate SITRANS LR250
•
increase measurement response if possible
•
clean the antenna
•
use Auto False Echo Suppression (2.8.7.1.) and Auto
•
False Echo Suppression Range (2.8.7.2.)
decrease Near Range (2.8.1.) (minimum value depends
•
on antenna type)
raise SITRANS LR250
•
ensure Algorithm (2.8.4.1.) is set to F (First echo)
•
ensure Material (2.2.3.) selection is LIQUID LOW DK
•
set Position Detect (2.8.4.2.) to Hybrid
•
set CLEF Range (2.8.4.4.) to 0.5 m
•
Diagnosing and troubleshooting
10.4 Operation troubleshooting