Auto Calibration
Please make sure the monitor is in
dark room,and the sensor probe is
put in the center marker and close
to the screen.
Select MENU to start!
MENU:OK
5、 Select " YES " to begin auto calibration. The color calibration instrument will automatically mea
sure the color of the screen and correct the color of the screen. During this process, it is necessary
to observe the color calibration progress bar in the color correction prompt box.
Auto Calibration
Progress
23%
Please wait...
Select INFO to Cancel!
6、 Press "Info/Quit" to terminate the color correction process at any
time. When the prompt color calibration progress reaches 100%,
the whole automatic color calibration is completed. After auto-
matic color correction, press the " Info/Quit " button to exit the
menu and let the monitor enter the normal display mode.
7、 After the automatic color correction, the display screen pops
up "Before Calibration" and "After Calibration" .
*4. Measure
The monitor has been calibrated in factory. And may need to
be re-calibrated after a period of time.
Before re-calibrated, the measure function can check the
current color to compare with the last time calibrated color,
to decide if the monitor needs to be re-calibrated.
Connect with the sensor probe and place the sensor probe
onto the right position like calibration step. Enter "Auto
Calibration" ‒ "Measure" .
The monitor will generate several colors and finish measure
within 30 seconds. And display the result as:
INFO:Quit
INFO:Quit
Auto Calibration
Please make sure the monitor is in
dark room,and the sensor probe is
put in the center marker and close
to the screen.
Select MENU to start!
MENU:OK
INFO:Quit
Before Calibration
After Calibration
R
R
R75
R75
Mg
Mg
Mg75
Mg75
B
B
B75
B75
Cy
Cy
Cy75
Cy75
G
G
G75
G75
YI
YI
YI75
YI75
W
W
Delta = 0.9
Delta = 2.3
CIE and Delta E
CIE and Delta E
before calibration
after calibration
Last Calibration
New Measured
R
R
R75
R75
Mg
Mg
Mg75
Mg75
B
B
B75
B75
Cy
Cy
Cy75
Cy75
G
G
G75
G75
YI
YI
YI75
YI75
W
W
Delta = 0.9
Delta = 2.3
CIE and Delta E
CIE and Delta E
form last calibration
form new measured
20

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents