X-Ray Room Requirement - Kodak 9000 Installation Manual

3d extraoral imaging system
Hide thumbs Also See for 9000:
Table of Contents

Advertisement

X-Ray Room Requirement

IMPORTANT
Use an appropriate wall fixing system suitable for the type of wall
construction. See the examples below.
The following illustrations provide examples of wall types and fixations.
2 x
16 mm
2 x
Table 4–3
X-Ray Room Requirements
Room Components
Width of the door
Height of the ceiling
Strength of the wall
Load-bearing capacity of
the floor
Required space for the
KODAK 9000 3D Unit
IMPORTANT
Install the Unit where a minimum amount of space is provided to allow
easy access for the patient or the maintenance technician (see Figure 4-2).
16 mm
Minimum Requirement
75 cm (30")
240 cm (95")
Withstanding an extraction force of 150 kPa
at each point of attachment.
500 kg/m2
1500 (L) x 2000 (D) x 2400 (H) mm
KODAK 9000 3D Extraoral Imaging System_Installation Guide (SM711)_Ed 01
2 x
16 mm
2 x
16 mm
Comments
If needed, a shorter column is possible
by special order.
It is the responsibility of the installer to
choose an appropriate fixing system that
withstands the extraction force.
The floor must be solid and flat.
X-Ray Room Requirement
4–7