Kodak 9000 Installation Manual page 28

3d extraoral imaging system
Hide thumbs Also See for 9000:
Table of Contents

Advertisement

Electrical Requirements
4–6
SITE PREPARATION BEFORE INSTALLATION
Table 4–2
Electrical Installation Specifications
Contactor (13)
Door Safety Switch
(9)
Use the following switches:
Emergency Stop Switch:
Manufacturer:
Ref:
Contactor:
Manufacturer:
Ref: LC1 D4011
IMPORTANT
If you cannot use the above manufacturer devices, use an equivalent
emergency switch and contactor in the UL list with the same specifications.
For more information, see UL 2601.1 clause 22.7.
16A-250V
20A-250V
UL listed
UL listed
1A / 250V
1A / 130V Optionally, connect the door safety
Cutler-Hammer
Telemecanique
switch (9) that deactivates the x-ray
remote control if the door remains open.
E22AT111
E22AT112
F6 (110/130 V - 60 Hz)
P5 (230/240 V - 50 Hz)
U6 (230/240 V - 60 Hz)

Advertisement