Kodak 9000 Installation Manual page 50

3d extraoral imaging system
Hide thumbs Also See for 9000:
Table of Contents

Advertisement

Installing the Unit
6–10
INSTALLING THE UNIT
line filter.
Connect cable 23 to J9 (D) of the CJ653 board
Connect the cables on the fixed arm to the head in the following order:
Connect cable 5 to J6 (E) of the CJ699 board.
Connect cable 1 to J3 (F) of the CJ732 board.
Connect cable 4 to J19 (G) of the CJ732 board.
Connect the cables on the left side of the column to the head in the following order:
Connect cable 12 to J8 (H) of the CJ732 board.
Connect cable 36 to J5 (J) of the CJ699 board.
Connect the Ethernet cable (K).
Connect cable 37 to J8 (L) of the CJ699 board.
The Unit is ready for the post-installation control.