Tests and Adjustments
Adjustments
Procedure
1. Connect the equipment as shown in
between the network analyzer and the power meter.
Figure 3-15
Equipment Setup for the Source Adjustment
2. .Press UTILITY
Source Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.

IF Gain Adjustment

The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1. Press UTILITY
Gain Adjustment.
2. Follow the instructions and prompts as they are displayed.

Receiver Characterization

This characterizes the receivers in your analyzer.
Procedure
1. Press UTILITY

Receiver Characterization.