10 Restart the computer, turn on both AA instruments and then
The instruments are now ready to be used as a simultaneous
flame/furnace system. Refer to the Help for further information on
dual instrument operation.
During dual instrument operation do not turn any connected instrument ON or
NOTE
OFF while readings are being taken.

Starting the SpectrAA Software

To start the SpectrAA software:
1
2
Ensure the software registration is completed by the user of the SpectrAA
NOTE
software. For further information refer to the Software Registration Help.
3
56
In the Run dialog, for Windows 7 type C:\Program Files
(x86)\Agilent\sp100\Run\DUAL.EXE or for Windows XP type
C:\SP100\RUN\DUAL.EXE and then select OK to start the
program.
A window displays asking you to turn on the instrument you
wish to reconfigure. You must select 'Instrument 2' in the
configuration. When you have switched on the instrument, select
OK.
A command window is displayed. The process may take about a
minute. During this time you will hear the instrument resetting.
Restart the computer and open the SpectrAA. As soon as the
Index page is displayed, shut down the PC again. This
automatically creates the necessary configuration files.
start the Agilent SpectrAA software.
Click Start > (All) Programs > SpectrAA and then SpectrAA
again. Alternatively, double-click the SpectrAA icon on the
desktop.
The first time the SpectrAA software is opened a Software
Registration dialog will appear. Click Next.
Complete all the fields on the 'Customer Details' page. Click
Next.