a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the WatchPAT is used exceeds the applicable RF compliance level above, the
WatchPAT should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the WatchPAT.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
WatchPAT™ONE System
TABLE 3 - FROM IEC 60601-1-2:2014