An estimation of the error in each snoring level was calculated by looking at the WatchPAT™
device measurement cut by the results of dB-meter in intervals of 1 dB in the range of above
40dB (below 40 dB was considered not clinically significant being background noise). A high
correlation was observed between the results of the two devices for the range of 40-70dB
(where sufficient data points were gathered), meaning the resemblance in the results
uniformly existed for all the snore levels measured.
The next table presents the statistics of WatchPAT™ device measurements per dB-meter
calculation at that range.
PSG
DB
N
Mean
Value
40
2033
41.10
41
1319
41.61
42
908
42.68
43
746
44.12
44
719
44.75
45
643
45.90
46
602
46.45
47
590
47.39
48
568
48.03
49
414
48.56
WatchPAT™ONE System
Coef. Of
Std
Variation
[%]
1.89
4.60
2.67
6.43
3.79
8.88
4.49
10.19
4.65
10.39
5.07
11.04
5.17
11.13
5.31
11.21
5.17
10.76
5.33
10.97
54
Min
Max
Median
40
54
40
40
54
41
40
62
41
40
58
42
40
65
43
40
59
45
40
59
46
40
66
47
40
61
49
40
64
49
Lower
Upper
95% CI
95% CI
41.01
41.18
41.47
41.76
42.44
42.93
43.80
44.44
44.41
45.09
45.51
46.30
46.04
46.86
46.96
47.82
47.60
48.45
48.05
49.08