Guidance and manufacturer's declaration – electromagnetic immunity – WatchPAT
The WatchPAT is intended for use in the electromagnetic environment specified below; The customer
or the user of the WatchPAT should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
WatchPAT™ONE System
TABLE 2 - FROM IEC 60601-1-2:2014
IEC 60601-1-2
Compliance level
Test level
+-8 kV contact
Not Applicable
+-2 kV, +4 kV,
+-2 kV, +4 kV, + 8
+ 8 kV, +15
kV, +15 kV air
kV air
30 A/m
30 A/m
Electromagnetic environment -
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30% applicable.
Power frequency magnetic fields should
be at levels characteristic of a typical
public low-voltage power supply network
that supplies buildings used for domestic
purposes, commercial or hospital, clinic
environment.
61
guidance