Itamar Medical OM2196470 Operation Manual page 27

Table of Contents

Advertisement

Itamar Medical Ltd.
NOTE:
Insert the battery into the device just before bed time, so it is full when the test
starts.
Visually inspect the battery before insertion, to ensure it is not swollen, cracked,
leaking or has other defect.
Notes/ Conditions for Battery Use:
1. The recording durations depend on the available life time of battery. It is important to insert
the battery just before usage.
2. The battery will be checked during device self-test, the WatchPAT will notify the patient in
case the battery power is low.
3. If battery was improperly inserted or is depleted the WatchPAT will not turn on. In this
unlikely situation, the patient should replace the faulty battery with a new AAA Alkaline battery,
purchased in a local store.
4. Battery should not be stored in the WatchPAT battery compartment, and only be inserted
when the patient is ready for the night test.
3.4.2
Wearing the WatchPAT™ONE device
The WatchPAT components are to be applied on specific location that will provide the required
signals. The sensors should be applied on the –
1. Wrist
2. Finger
3. Chest (if a chest sensor is provided)
3.4.3
Strapping the Wrist Device
The first step would be applying the wrist device. Place the wrist strap on the non-dominant
arm and close it snugly but not tightly (see Figure 5). Ensure that the side connected to the
Finger Probe is towards the fingers.
You may find it convenient to place the wrist strap with the WatchPAT device facing down on
the table and then place the back of the wrist over the wrist strap in order to fasten the straps.
WARNING
Do not close wrist strap too tightly.
WatchPAT™ONE System
26
Operation Manual