Itamar Medical OM2196470 Operation Manual page 50

Table of Contents

Advertisement

Itamar Medical Ltd.
Error Message
Device is not located. Please
check if WatchPAT ONE's LED
blinks. If it does, place your
phone closer to the device.
If not: verify that you placed
a new battery and check it is
properly positioned
Device battery is low or the
device's battery is depleted or
damaged. Please replace the
battery and try again.
Multiple devices are identified
in the surrounding. Please
remove battery from all
irrelevant devices and try
again.
The WatchPAT is asking to turn
on Bluetooth
WatchPAT™ONE System
Possible Reason
The Application cannot find an
active device in its proximity
The device battery has run out
of power
The Application sees more than
one active device
The Mobile phone does not have
its Bluetooth communication
turned ON.
49
Action
If there is no blinking on the
device cover, check if the
battery in the device was
properly placed and press NEXT.
If it is blinking, bring the device
closer to the Phone and press
NEXT.
Check if the Bluetooth is
enabled on your device. If there
is still no connection created,
pull the device's battery out,
select Forget Device from menu
and start again.
If there is still no connection
created, try closing other
Applications that use the BLE.
If the RED blinking still resumes,
there is a problem with the
device, and it needs to be
returned.
Replace the battery with a fresh
Alkaline AAA battery
You should make sure other
WatchPAT ONE devices in
room are turned OFF (batteries
removed) until after your
Application has successfully
establishes communication
with your device
Approve the Application's
request of the turning ON the
BlueTooth capability.
Operation Manual