Itamar Medical Ltd.
Guidance and manufacturer's declaration – electromagnetic emissions – WatchPAT
The WatchPAT is intended for use in the electromagnetic environment specified below; The customer
or the user of the WatchPAT should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
WatchPAT™ONE System
TABLE 1 – FROM IEC 60601-1-2:2014
Compliance
The WatchPAT uses RF energy for internal function
and for BLE (4dbm) transmissions. Therefore, its RF
Group 1
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The WatchPAT is suitable for use in all
Not
establishments, including domestic establishments
applicable
and those directly connected to the public
low-voltage power supply network that supplies
Not
buildings used for domestic purposes.
applicable