Checking The Scale Settings Concerning Scale Verification; Menu Overview In The Service Mode (Adjustment Switch In The Adjustment Position) - KERN MPS200K100NM Operating Instructions Manual

Personal weighing scale, handrail scale, bariatric scale, wheelchair scale transportation stretcher balance
Table of Contents

Advertisement

Available languages

Available languages

Position of adjustment switch:
Position of adjustment switch
1. left side
2. centre
3. right side

13.3 Checking the scale settings concerning scale verification

To start the adjustment function, a scale is to be switched over into the service
mode. To do this, set the adjustment switch in the adjustment position (refer to
chapter 12.2).
The service mode makes it possible to change all parameters of a scale. Service
parameters cannot be changed because it may have an influence on the scale
settings.
13.3.1 Menu overview in the service mode (adjustment switch in the adjustment
position)
Overview is only used to check the set parameters by authorised verification offices.
Changes may only be introduced in parameters of the automatic switching off
function „A.OFF" and audible signal „BURR".
MPS / MTS / MXS / MWS-BA-e-1621
1
2
Status
not documented
adjustment position - adjustment is
possible
verification position – adjustment locking
3
59