8.4 Adjustment using the external weight

• Factory setting in ALS/PLS models
• In ALJ/PLJ models it is available only in the verifiable setting.
• Preliminary
chapter 8.1.
• The weight of the recommended adjustment weight, see chapter 1
"Technical Specification".
• Information on reference weights can be found on the Internet at
http://www.kern-sohn.com.
 Ensure there are no objects on the scale plate. In the weighing mode, press the
CAL button.
 Wait until the blinking weight value of the required adjustment weight is
displayed.
 During blinking, place the required adjustment weight carefully in the middle of
the scale plate. The blinking symbol will disappear.
 After the successful adjustment, the scale will switch to the weighing mode again
automatically.
 Remove the adjustment weight.
,
38