Adjustment Report Display/Print - KERN ALJ Operating Instruction

Hide thumbs Also See for ALJ:
Table of Contents

Advertisement

Available languages

Available languages

8.6 Adjustment report display/print

This function enables to print the report on the most recent adjustment.
The scale and printer communication parameters must be compatible.
Communication parameters, see chapter 11.2 and 11.3.
Printout conforming to GLP see chapter 12.8.
 In the weighing mode press and hold the MENU button until the sound signal
stops. The configuration menu is displayed.
 Using the navigation buttons , select the menu item <Calibration info>.
 Confirm, pressing the PRINT button. The following will be displayed: date, time,
adjustment type and deviation when compared to the most recent adjustment.
 After the optional printer is connected, the data can be printed by pressing the
PRINT button.
Sample printout (KERN YKB-01N):
 Press the ON/OFF button. The scale will be switched to the menu again.
Whenever required, carry out other settings in the menu or press the ON/OFF
button. The scale will switch to the weighing mode again.
40
27-08-20 10:41:17
Balance ID:
WI2000077
User ID
Miller
Project ID
KERN
Calibration mode
27-08-20 10:11:17 AM
Internal calib.
Corr.: 0.21 g
Signature:
TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2112
Current date/time
Adjustment date/time
Adjustment type
Deviation when compared
to the most recent adjustment

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents