General information for the following adjustments

1. Switch on the ARCADIS system.
2. In the application menu bar, click on <Options>-<Service>-<Local Service>.
3. Enter the (6-digit) password.
4. Click on <OK>.
5. Click on <Main System>.
6. Select the ARCADIS system type.
7. Click on <Next>.
The following adjustments are available:
Generator adjustment
Generator adjustment check
Load counter
Image rotation
Collimator X-iris
Display X-iris
Collimator slot diaphragm
Display slot diaphragm
Dose area product
Air kerma
Dose rate adjustment / checking the dose rate in automatic mode
These adjustments are described on the following pages in the above-mentioned
sequence in this document, and they are also available in the system. A <click here> link
is available on the pages where adjustments are possible. When you click on <click
here>; the relevant adjustment description page will come up on the monitor. Always
place the description page on the right monitor.
NOTE
NOTE
ARCADIS Varic / Orbic
After making changes under "Configuration" or "Adjustment", a
backup of the new values is required.
After you enter the service menu, the currently selected dominant
is shown as a circle in the radiation field.
SPR2-310.842.01.07.02
12.06
Page 6 of 34