Adjustments
4. Enter the chamber constant from the monitor acquisition task card under <Main Sys-
tem>-<Configuration>-<Option> "DAP chamber resolution".
5. Click on <Finish>.
6. Continue with "DAP accuracy check" or "Air Kerma accuracy check".
PTW - chamber test and adjustment
1. Read out the TW constant from the previous page or under <Main System>-<Configu-
ration>-<Options>.
2. Click on <Go> to start the DAP test.
3. Read out the DAP constant from the ARCADIS monitor at the acquisition task card.
4. Click on <Go> again to stop the DAP test.
5. Compare these two values.
Fig. 6:
Siemens AG
Medical Solutions
With "DAP" display: Multiply the monitor value by 100 and enter the 4 digits (e. g.
49.97 x 100 = 4997).
With "Air Kerma" display: Enter the first 4 digits from the monitor value. (e. g.
4997.2 = 4997).
Status bar: success
Status bar: success
Window appears: Calibration is finished --> confirm with <OK>.
The main system restarts.
The DAP constant TW and the value from the ARCADIS monitor acquisition task
card should be equal.
Dose Area Product Adjustment
SPR2-310.842.01.07.02
12.06
Page 23 of 34
CS SD 24
23
ARCADIS Varic / Orbic