Emc Emissions And Immunity - MAGSTIM RAPID2 Operating Manual

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

2
Magstim Rapid
SECTION 11: EMC EMISSIONS AND IMMUNITY
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
2
The Magstim Rapid
System is intended for use in the electromagnetic environment specified below. The
customer or the user of the equipment should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
EN60601-1-2
RF emissions
Class A
EN60601-1-2
Harmonic
Class A
Emissions
(by equipment
EN60601-1-2
type)
Voltage
fluctuations/ flicker
Complies
emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
2
The Magstim Rapid
System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Magstim Rapid
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Magstim Rapid
output power of the communications.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
The MAGSTIM Company LTD
Electromagnetic environment - guidance
The Magstim Rapid
order to perform its intended function. Nearby electronic equipment
may be affected.
The Magstim Rapid
other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
2
System as recommended below, according to the maximum
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
3,5
]  P
d = [
d = [
V
1
Where V
= 3
Where E
1
0,117
0,369
1,167
3,689
11,667
59
2
System must emit electromagnetic energy in
2
System is suitable for use in all establishments
2
System can help prevent
m
800 MHz to 2,5 GHz
3,5
]  P
d = [
E
1
= 3
Where E
1
0,117
0,369
1,167
3,689
11,667
MOP03-EN-01
7
]  P
E
1
= 3
1
0,233
0,738
2,333
7,379
23.333
June 2011