Isolated Trigger Port; Sd Card File Details - MAGSTIM RAPID2 Operating Manual

Hide thumbs Also See for RAPID2:
Table of Contents

Advertisement

2
Magstim Rapid
7.2 ISOLATED TRIGGER I/O PORT (1)
WARNING
Pin Connections.

7.3 SD CARD FILE DETAILS

General
Each line of the file starts with a text string that describes the data that follows on that line.
Each line of data ends with a carriage return line feed request, <CR>.
File Layout
The following details the order the data is stored in the Magstim UI generated file.
Version, <file format number>, <CR>
File, <system filename>, <CR>
Status, <system file status when exported>, <CR>
File Time, <seconds>, <minutes>, <hours>, <day>, <month>, <year>, <CR>
Name, <patient name>, <CR>
Sex, <patient sex>, <CR>
ID#, <patient ID>, <CR>
Date or birth, <patient date of birth>, <CR>
Protocol description, <patient protocol description text string>, <CR>
Comments, <patient comments string>, <CR>
Train, <power>, <frequency>, <duration>, <number of pulses>, <wait time>, <number of trains>,
<CR>
Burst, <power>, <frequency>, <number of pulses>, <burst frequency>, <number of bursts>, <cycle
time>, <number of cycles>, <CR> *see note1.
Session 1, <start time>, <power>, <frequency>, <duration>, <number of pulses>, <wait time>, <CR>
Session 2, <start time>, <power>, <frequency>, <duration>, <number of pulses>, <wait time>, <CR>
.
.
.
.
Session 30, <start time>, <power>, <frequency>, <duration>, <number of pulses>, <wait time>, <CR>
MEP 0, Trace 1,Trace 2, <CR>
MEP 1, <reading 1 trace1>, <reading 1 trace2>, <CR>
MEP 2, <reading 2 trace1>, <reading 2 trace2>, <CR>
MEP 3, <reading 3 trace1>, <reading 3 trace2>, <CR>
MEP 4, <reading 4 trace1>, <reading 4 trace2>, <CR>
.
.
.
.
MEP 300, <reading 300 trace1>, <reading 300 trace2>, <CR>
MEP Configuration, <time base index>, <channel 1 voltage range index>, <channel 2 voltage range
index>, <special feature>, <filters index>, <CR>
The MAGSTIM Company LTD
Only equipment complying with EN 60601-1, and configured in compliance
with EN 60601-1, should be connected to the trigger output and/or trigger
input. Connection must be via Trigger Input/ Output Cable P/N 600003 in
order to maintain System EMC Compliance
I/O – 5V CMOS Logic Levels.
1. Aux GND
2. N/C
3. Tout- O/C
4. N/C
5. +ive Trigger In
6. -ive Trigger In
7. +ive Trigger Out
8. -ive Trigger Out
9. N/C
10. Aux +5V
11. Aux GND
12. N/C
13. N/C
.
.
.
.
.
.
.
.
.
.
.
.
43
14. N/C
15. N/C
16. N/C
17. N/C
18. N/C
19. Aux GND
20. End Train
21. N/C
22. N/C
23. N/C
24. Trigger Edge
25. N/C
26. N/C
.
.
.
.
.
.
MOP03-EN-01
.
.
.
.
June 2011

Hide quick links: