BACKUP BATTERIES

9.4 CLEANING AND DISINFECTING

9.5 SERVICING

The MAGSTIM Company LTD
Battery status can be checked on the user configuration screen and when
low, is reported on the main Options Menu screen.
The batteries in the UI will require to be replaced every 2 years. To do so,
unplug the UI connector, turn the UI upside down, remove the 4 cross-head
screws visible, and detach the plastic bottom of the UI. Have the new
batteries available. The UI has a small capacitor which will maintain the
internal settings for about 20 seconds once the old batteries are removed.
It is important that the batteries are replaced within the 20 seconds if data
is not to be lost. Before removing the old batteries, ensure you have the
correct replacements, and that you are clear about the orientation of the
batteries. Once inserted, replace the bottom of the UI and fix using the
screws. Do not over-tighten.
Do not use re-chargeable batteries.
When spent, the batteries should be disposed of under the appropriate
environmental regulations
It is advisable that the batteries be removed from the UI if the equipment is
likely to be un-used for a long period of time.
The stimulating coil, exterior of main frame, PSU, UI, foot switch, MEP
Pod and MEP Pod patient cables may be cleaned using an isopropyl
alcohol moistened cloth. Ensure that the equipment has dried thoroughly
before use.
Note 1: The coil cannot be sterilised. Do not allow the coil to become
contaminated with body fluids.
Note 2: Do not clean or wipe the touch screen with anything abrasive as it
will cause permanent damage.
All servicing must be carried out by the Magstim Company Ltd, or one of
its authorised service centres.
It is recommended that an annual PAT (Portable Appliance Test) test be
carried out on the Magstim Rapid
For further information or service contact:-
The Magstim Company Limited
Spring Gardens
Whitland
Carmarthenshire
SA34 0HR
Wales
United Kingdom
Tel: +44 (0) 1994 242900
Fax: +44 (0) 1994 242910
email: service@magstim.com
Website: www.magstim.com
2
Important: The Rapid
, PSU, Plus
mains/ HV potential during operation. Some HV/ hazardous potentials will
continue to be present for some time after the system has stopped
operating. Therefore, under no circumstances must un-authorised persons
remove the screws securing the covers of the Magstim Rapid
49
2
.
1
and UI contain circuitry that is at
MOP03-EN-01
2
1
, PSU, Plus
June 2011

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents