4.7 REPETITIVE MODE: BURST SETUP
Select Burst Mode from the Repetitive Setup screen to allow the following adjustable parameters to be
set.
If Repetitive Setup is accessed from Single Pulse Mode, the protocol intensity value established in
Single Pulse Mode will be carried over into the power window.
Selecting the Power window allows you to adjust the power level between 0%
Power
and 100% using the Rotary Control Knob.
Frequency
Selecting the Frequency window will allow the user to set the frequency of the
pulses to a value between 0 and F
frequency can be up to 100Hz depending on the power level set and the
system configuration. A table detailing the available power/ frequency
combinations is included in Section 10.
Selecting the „Number of Pulses‟ window allows the user to set the number of
Number of
Pulses
pulses per Burst. The Number of Pulses can be set between 1 and 10 using the
Rotary Control Knob.
Number of
Selecting the „Number of Bursts‟ window allows the user to set the number of
Bursts
bursts in a cycle. The number can be set between 1 and 500 in increments of 1.
Selecting the „Burst Frequency‟ window allows the user to set the Burst
Burst Frequency
Burst Frequency.
frequency between 1Hz and 10Hz.
Selecting the „Cycle Time‟ window allows the user to set the time for each
Cycle Time
cycle in seconds. The time can be set between 1 and 500 seconds in increments
of 0.1 seconds. However, if the minimum cycle time required for the selected
protocol is greater than that selected, the UI will display a message „Cycle time
too short, auto-set to minimum‟. Select „YES‟ to set the cycle time to the
minimum allowed for the chosen protocol. Select „NO‟ to reset the protocol
parameters manually.
Selecting the „Number of Cycles‟ window allows the user to set the number of
Number of
cycles in a session. The value may be set between 1 and 999 in increments of 1.
Cycles
The number of cycles cannot be set to exceed the maximum total number of
.
pulses of 65000.
Print
For Print operations See Section 6.5.
Run Session
This screen takes the user to the Repetitive Mode Screen to run a session.
Save Session
Save Session takes the user to the File Maintenance screen (See section 6.1).
Options Menu
Selection of this option will return the user to the Options Menu screen.