9.2 USER MAINTENANCE

9.3 TECHNICAL MAINTENANCE

The MAGSTIM Company LTD
At the start of each session the Operator must check the Magstim Rapid2,
PSU, UI and the stimulating coil for any signs of damage, paying particular
attention to the plastic casing. If there are any signs of physical damage to
the Magnetic Stimulator, PSU, UI or the stimulating coil, they must not be
used.
The cooling vents should also be checked to ensure that they have not
become blocked, or obstructed.
The UI contains 2 AA cell batteries. These are not rechargeable, and will
need to be replaced if patient information is stored on the system. It is
important that regular checks are carried out to ensure that the batteries are
replaced when necessary. Failure to do so will result in the information
being lost. The Battery voltage can be viewed in the User Configuration
Screen.
The coil pins should be checked before each session for any signs of pitting,
or burning, as under conditions of exceptionally hard use at high energy
levels, it is possible for the localised heating to manifest itself in the form
of micro-welds. Continued used in this condition will eventually result in
the coil pins/sockets becoming totally eroded and open circuit.
Note: The pin burning is communicable - any coil with good pins that are
used on a stimulator with burned socket pins will have its pins damaged
immediately. The reverse is also true; a socket with good pins that has a
coil with burnt pins connected to it will have its pins damaged
immediately. If damage is noticed on any coil connector or stimulator coil
socket the complete system must not be used until all pins and sockets are
carefully examined for any damage. If any contacts show damage, even if
slight, they will need to be changed. If this is not done thoroughly there is
the risk that the cycle of contact damage will continue.
As contact repair is a specialised procedure, it is recommended that contact
replacement is undertaken by the Magstim Company, or one of its
authorised service centres. If additional advice or information is required,
please contact The Magstim Company.
48
MOP03-EN-01
June 2011