This is useful for obtaining printed verification of load
cell accuracy to comply with ISO-9000 and British Standards
rules.
The recommended procedure is:
1.
Place the unit into the Program mode.
2.
Press *88 for printout of empty bin TARE weight.
3.
Place a KNOWN CERTIFIED WEIGHT into the weight bin.
4.
Press *88 again for printout with the weight added.
5.
The different between the two weight printouts should
equal the KNOWN CERTIFIED WEIGHT.
*89
Press (*,8,9) to select the desired weight unit (GRAMS,
KILOGRAMS, OUNCES, POUNDS) for readout of data.
the U.S.A. are supplied with POUND readout selected.
KILOGRAMS is preferred for nearly all countries outside the
U.S.A.
short runs or short demonstrations.
Use the * key to walk through the four possible selections.
Press EXIT when the weight unit you want is displayed.
*98
Press (*,9,8) to set flag for RAW SIGNAL readout in place
of gram readout of scale weight.
useful to demonstrate the extreme sensitivity of the load
cells.
Load cell function can be monitored without concern for any
improper weight calibration that may have been done.
Use the * key to toggle flag ON or OFF.
Power On will always set this flag to OFF.
Press EXIT when desired flag status is displayed.
During normal operation, pressing the "CE" key will cause
raw signal readout to be displayed for 5 seconds.
Date:
Time:
Machine number:
Display Readout:
GRAM or OUNCE readout is appropriate only for very
The raw signal readout bypasses calibration math.
Page 43
11/09/93
17:22:01
002
P
500.0