8. Calculate the amplitude loss with the following formula: Amplitude Loss (dB) = 20
x lg (Vrms2/Vrms1).
9. Repeat Step 2-7. Measure Vrms1 and Vrms2, and calculate the amplitude loss of
500 mV/div vertical scale.
10. Turn off CH1. Measure CH2, CH3, and CH4 respectively using the method above
and record the measurement results.
5.6.4

Test Record From

1 MΩ Input Impedance
Channel
CH1
CH2
CH3
CH4
50 Ω Input Impedance
Channel
CH1
CH2
CH3
CH4
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Vertical
Scale
V
1
rms
50 mV/div
500 mV/div
50 mV/div
50 mV/div
50 mV/div
500 mV/div
50 mV/div
500 mV/div
Vertical
Scale
V
1
rms
100 mV/div
500 mV/div
100 mV/div
50 mV/div
100 mV/div
500 mV/div
100 mV/div
Performance Verification Test
Test Result
Amplitude
V
2
rms
[1]
Loss
Test Result
Amplitude
V
2
rms
[1]
Loss
DHO4000 Performance Verification
Limit
Pass/Fail
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
Limit
Pass/Fail
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
-3 dB to 3
dB
31